INSTRUCTIONS TO FIX LEAK IN VENOUS PORT OF CALIFIA PATIENT MODULE (CPM)
BACKGROUND:
Use these instructions to service a leak at the junction between the venous panel-mount quick connector and the venous valve threaded white bushing.
- PREPARATION:
- Drain the reservoir in its entirety
- Power off the unit
- Disconnect all external tubing
- Disconnect the power cable and Ethernet cable
- Move the Califia Patient Module (CPM) to a table or benchtop with plenty of workspaces
- REQUIRED TOOLS

Table of Contents