BACKGROUND:
The level sensor used in a Califia Patient Module reservoir measures fluid volume by sensing electrical capacitance inside the reservoir. The following instructions can be used to configure the level sensor.
- PREPARATION:
- Preparing Califia Patient Module (CPM) for level sensor configuration
- Connect the power cable and ethernet cable to CPM
- Connect circuit tubing (Arterial, Venous, Cardioplegia, and Vent)
- Power up the CPM
- Launch the Califia Simulator software; it should recognize the CPM
- Preparing Califia Patient Module (CPM) for level sensor configuration
- CONFIGURATION:
- Level sensor teaching process
- In Opening Screen, select CPB and press the Continue button
- Press the Califia Controls button, ① in Fig 1
- Level sensor teaching process

Figure 1. Califia Controls settings before level sensor teaching process
- Press the Open Valve button to disable level control, ② in Fig 1
- Set the Venous Valve slider to 100%, slider below ② in Fig 1
- Press the Calibration Panel button, ③ in Fig 1

Figure 2. Calibration settings before level sensor teaching process
- Set the Calibrate slider to Level, ① in Fig 2
- Press the F11 key on the keyboard to bring up the level teach controls, see ② in Fig 2
- Clamp the venous return and use a pump to fill Califia’s reservoir to the 700 mL mark
- Once a volume is stable (not moving around), press the Low-Level Teach button, ① in Fig 3. The process takes about 20 seconds to complete. You will hear a series of audible clicks coming from the Califia Patient Module computer board. The indicator labeled Sensor (1L) should read around 0.40.
Note: if Sensor (1L) reads significantly different than 0.40, repeat this step to teach low level.

Figure 3. Level Teach buttons in Calibration section
- Fill Califia’s reservoir to the 3.6 L mark
- Once a volume is stable (not moving around), press the High-Level Teach button, ② in Fig 3. The process takes about 20 seconds to complete. A few audible click sounds are made. The indicator labeled Sensor (3L) should read between 3.50 and 3.65.
Note: if Sensor (3L) reads outside the specified range, repeat this step to teach high level.
- Level sensor Calibration process
- Empty the Califia reservoir to the 1 L mark
- Once a volume is stable; press the checkmark box next to 1 Liter, ① in Fig 4
- Fill the Califia reservoir to the 3 L mark
- Once a volume is stable, press the checkmark box next to 3 Liters, ② in Fig 4, then press Apply button, ③ in Fig 4
- In the Califia Controls section, verify Level indicator reads 3 Liters; verify it reads properly at two other volume marks, say 2.5 L and 1.5 L.

Figure 4. Calibrate level
- Press the Valve Opened button to enable level control
- Test the Califia system by going on and off bypass to make sure the system responds as expected